RefleX Support -Release Notes

1.00  1.01  1.02  1.03  1.04  1.05  1.06  1.07  1.08 1.09  2.00 2.1 2.2 2.3 2.4 2.5 2.5.1 2.6

1.00

1.)  Initial Release                                                                                  * = enhancement)

1.01

1.) *Foreign language implementation.  Instructions
2.)  Calculation of DEV angle is incorrect in True Position mmc

1.02

1.) *Service tool modifications (allow weaker encoders)
2.)  Point zeroing in only one axis as per manual.
3.)  Tolerancing arrow is a down arrow: consistent with other field movement operations.
4.)  In HT GAGE Mode, it is possible to report both the incremental and the
      absolute distance.
5.)  Export of statistics data in all modes.

1.03

1.)  Checksum in flash card attribute memory
2.)  AutoTune improvements.

1.04

1.)  Translation of datum in playback.
2.)  3D distance of point to origin.
3.)  Translate datum by part vs. machine coordinates.
4.)  Set origin to circle/cylinder change level.
5.)  Qualify tapered/ball/edge/optical probe after qualification of TTP does not work in
      playback.
6.)  Rotate datum in playback.
7.)  Spheres outside only.
8.)  Rotate datum wrong direction.  (Pos. should be CCW)
9.)  Polar coordinates angle not +/-   (point, circle, cylinder, sphere, cone).
10.)  Z-Mouse delay 500/200.

1.05 (5/5/97)

1.) *Dim construction button until valid features have been measured.  Lines and planes
       are not used in constructions.
2.) *Make scroll keys repeat when held down in program list and playback go-to-block
       list.
3.) *Screen numbering now reflects number of screens available to the user instead of
       number of possible screens.
4.) *AutoTune to read EEPOT values from PROM.
5.) *Remove "Adjust Qualification Sphere" after first tip measured.
6.) *Recall probe tip to default to current tip.
7.) *Change system to display xxxx.xxx(mm) and xxx.xxxxx(in).
8.) *Change "Volcomps do not match" screen to say "controller" instead of "machine" to
       avoid confusion.
9.) *Dim buttons while printing screens.  Customer thinks he can press them and he can't.
10.) *True Position diameter Tolerancing to include upper and lower instead of just total.
11.) *Dim part drawing button if no features measured.
12.) *Print Speed option in Measurement setup is not being saved in PROM.
13.) *Rene LaForest- Reword "Reset Measurement Mode".  Currently looks like an
         option list. Add text:  "The following items will be reset:"
14.) *Autotune:  Fix overflowing bar graphs in align, implement new method to read
         hardware to get around bug in chip, return new error codes.
15.) *Program creation measured values, instead of entered nominal values, are
         being shown in the playback prompt "Measure this feature" screen.
16.) *TTP taking hits in the startup menu before in a measuring mode.  Fixed for
         all screens!
17.) *Store all variables in 8k EEROM to fix boards without 2k chip. This will fix all
         controllers which are losing their system parameters!!!
18.) X and Y for relationship between a point and a line are a probe diameter off.
19.) Preserve user alignment and tip when saving a program to the card.
20.) Enable serial output in Height Gauge mode and XYZ Counter mode.
21.) Clear existing reference feature when "Reset measurement mode" button pressed.
22.) Print concentricity and automatically redraw the list in output selection menu
       whenever the printer is enabled or disabled.
23.) Fix save datum in playback to function correctly.
24.) Fix True Position Printouts.  Add header to TP-tolerance on LCD.
25.) Correct playback from DRAM and PCMCIA card so that if a reference feature
       is invoked, the relationship is computed with the reference feature and not the
       previous feature.
26.) Correct software to give consistent results for squareness and parallelism when using
       relationship. (Fixes distance between two constructed line problem).
27.) Remove "Back and Forth" from AutoTune Screen.
28.) Comma Delimiter not displayed in TTP preview.  Comma delimited little
       counters overwrite X,Y,Z.
29.) Symmetry line has correct I, J, K but wrong X, Y, Z position.
30.) Symmetry plane has correct I, J, K but wrong X,Y,Z position. 
31.) Multitip qualification to dim "Next Tip" button after 9th tip measured.
32.) Tip #7 is always 0, Tip #8 is always two probe diameters.
33.) Out of order reference features are a problem in playback
34.) Cancel "Adjust Qual Sphere Dia" should abort to "Measure Qual Sphere Dia".
35.) Cancel button in qual of multiple tips should abort only current tip qual and not
       whole probe.
36.) Cancel "Measurement Setup" should abort to probe menu.
37.) Allow system to remember the chosen mode (text or graphic) for specific blocks to
       give desired output in playback.
38.) Give correct polar angles for points.  Polar coordinated compensated along radius
       in datum plane.
39.) Connect first and last points of circle form plot if within 30°.
40.) Gauge angle output being reported incorrectly.
41.) Removal of (Nominal/Upper Tolerance/Lower Tolerance) from Measured and
       None option list.
42.) VAREAS appear one by one after tolerances set to none.
43.) If measured value = tolerance max. then showing out of tolerance.  Should be
       in tolerance.
44.) Disappearing VAREAS ( X, and  Y) in True Position screen during playback.
45.) Double blinking cursors in playback.
46.) Retain all tip results for recall, even if a recall probe tip was deleted from a program.
47.) Problems with roundness screen in playback when only 3 points taken.
48.) Plot screen histograms for 3 point circles and 2 point lines are not being drawn
       correctly.
49.) Remove cancel button from "Can not solve feature" during probe qual in playback.
50.) While measuring qual sphere, if cancel is pressed, return back to "Measure
       Qual Sphere" screen.
51.) Parallelism and Squareness small values reverting to zero after being toleranced.
       Tol band is being incorrectly computed.
52.) When playback is finished printing a block, at the screen where it allows you to either
       cycle through screens, cancel out of playback, press done, cursor no longer blinks.
53.) True Position format will be set to output selection format unless previously modified.
54.) True Position screen to have TP an angle instead of dX and dY.
55.) Tip # removed from ball probe icon.
56.) Distinguish between Computed and User reference lengths.
57.) Line/Line Intersect using centroid of line instead of actual intersection point.
58.) System is not pausing for done in max. number of points screen.
59.) Sequence: Measurement setup help button help button again (to cancel out of help)
       cancel again (to cancel out of Measurement Setup)
60.) If a person rotates a datum, saves that datum, then decides
       to delete that block, he is still allowed to recall that datum.
61.) If reference feature selected just after a non-feature block (such as rotate datum)
       reference feature not relating to any features. Even after measuring more features.
       One must set reference feature again before it will work.
62.) After turning off reference feature (reference feature was previous feature) glist was
       17:17:18 instead of 16:17:18
63.) Perpendicular plane to line "Intersection Point and Gauge Angle" showing
       actual instead of gauge angle.
64.) In all angularity screens, angularity value not being recomputed after nominal
       angle value changed.
65.) Output setup (Text ((only) / Graphic (only) /Both) is reverting to text (only) on
       'Reset Measurement Mode' and rebooting.  Save output setup in PROM.
              Note:  Program playback should not affect this variable.
66.) Circles smaller than .5 cm are being solved as planes 90% of the time.
67.) Probe compensation of polar points being done incorrectly for certain directions.
68.) Canceling out of the first measuring screen of playback "scanning" screen displays
       "Measure Qualification Sphere" screen.
69.) Datum translation in playback:  When setting just on keyed-in value, other values
       change too.
70.) Playback at the end of probe qual:  Summary screen shows "learned" probe diameter
       instead of "measured" probe diameter
71.) Multitip probe qualification in playback summary screen shows only last 4 tips.
72.) Aborting out of playback (from PCMCIA) card only and scrolling up to unmeasured
       blocks displays garbage on screen.
73.) Aborting out of probe qual during "Select probe type" on start up puts user in
       main menu.
74.) At completion of playback, when you scroll back to tip qual in measurement mode,
       tip summary not being displayed.
75.) If reference feature same as last feature, don't show last block twice in glist17:17:18
76.) If straightness or roundness plot is the first screen of a block printed out, not printing
       block header above it.
77.) Sequence:  Measure a line with N number of points.  Print straightness (Block4).
       Set Datum 'B' (Block 5). Measure Line, turn on relationships. Print intersection
       (Block 6).  ==> Line(6) --> Roundness(4).  Should say Straightness(4).
78.) All angularity values are not being printed in Full style.
79.) Some squareness, parallelisms, and angularities are not being outputted to printer.
80.) Some parallelism, squareness, or angularity outputs are displaying tolerance value
       twice in full mode:  (Single line squareness), (Single plane squareness), (Single
       cylinder parallelism and squareness), (Single cone parallelism and squareness)
81.) Because Nominal = Actual, Some angularity values are being displayed as 0.000
       until toleranced
82.) Perpendicular line and cone 1st screen XY values bad:? Perpendicular line and cone
       2nd screen angle bad? Parallel line and cone 3rd screen perpendicular distance
       value bad?
83.) Line Line angularity not being computed correctly.
84.) Rotation angle of keyed in offsets does not seem to be working for negative values.
85.) (Circle cylinder) has right arrow highlighted even if will not show next screen.  Also
       in playback.
86.) Probe Qual in playback is sometimes quickly displaying an extra screen: "Insert
       Probe". Also displaying "1/0" in screen counter section of status bar.

1.06 (11/6/97)

1.)  *Don't restore global register G14 after interrupt as specified in errata.
2.)  *New additions to fault table as described in I960 errata.
3.)  *Change software to handle all interrupts instead of just timer interrupt.
         Increase size of interrupt stack
4.)  *Improved audible beep quality.  (Beep no longer executing in ISR).
5.)  *Larger print buffer allows user to continue measuring while printing out a
         part drawing.
6.)  *Lissajous Encoder Utility fully operational.
7.)  *Modifications to EEPOT driver result in much faster AutoTunes.
8.)  *PCMCIA storage card percentage full graphs.  Instructions
9.)  *Recall Alignments in Dial Indicator mode. Instructions
10.)  Controller Lockups Fixed.
11.)  Z-Mouse Lockups Fixed
12.)  Fix digitize lockup bug. System would 'hang' when entering digitize outline in
         certain situations.
13.)  Optical probe not able to measure line created with only 2 points.
14.)  Fix screen printouts so that there is not the gap between the bottom line and side
         lines of the screen perimeter.
15.)  Software not reverting to tip 1 on reset measurement mode as described
         in "Reset Measurement Mode screen.
16.)  Corrected value for include angle of perpendicular 'plane & cylinder' and
         'plane & cone'.
17.)  Change such that software no longer appears to lock when printing to an off-line
         Okidata printer.
18.)  True Position was always being reported in radians.
19.)  Constructed Projection Points now working correctly.
20.)  In measurement mode, Form Plot of plane in XY axis upside-down.
21.)  No longer deleting probe tips when jumping to first block from the
         "End of Playback" screen.
22.)  Reduced probe debounce time.
23.)  Datum translation in playback now working correctly.
24.)  Alignment problem in ZX plane fixed..

1.07 (4/1/98)

1.)  *Relationship between any two features.   Instructions
2.)  *Radius & Diameter option. Instructions
3.)  *Service Tool: "RefleX in a suitcase" demonstration
4.)  *Language Screen on/off. Instructions
5.)  *Central localization of all system options. Instructions
6.)  *User customizable decimal places  Inches = 5 or 4, MM = 3 or 2. Instructions
7.)  *Printable headers.  (User must fill in with pencil) Instructions
8.)  *Service Tool: "Autotune save on failure"
9.)  *Service Tool:  Machine parameters now based on machine type in System Options.
10.)  *User selectable cone angles- half or full. Instructions
11.)  *Capability to Qualify Additional Tip without having to requalify entire probe.
12.)  *Skip probe menu an go directly to probe qualification on startup.
13.)  *Improved Graphics for Probe Qualification and Volcomp Screens.
14.)  *Display support site web address at startup
15.)  *Undim Digitize Menu if no printer to avoid confusion.
16.)  *Individual printers listed in options so customers know which printers are
          supported
17.)  *Rate error screen now displays all possible causes of the error.
18.)  *Implementation of Swedish and Finnish language
19.)  *Allow Datum A for Lines and constructed lines.
20.)  *Serial output in inspect mode.
21.)  *Datum B for Cones and Cylinders and Lines.
22.)  2nd Intersection point between line and circle now being stored correctly.
23.)  True Position MMC bonus now displaying correct value.
24.)  Printer lockups appear to be fixed.
25.)  Show rotation angle when rotating by keyed in offsets to avoid confusion.
        Also, correctly show axes to be rotated about.
26.)  Don't show change feature button while in relationship mode.

 1.08 (11/10/98)

1.) *First Support for dash-6 controllers (Volume adj, calendar style Y2K clock).
2.) *Modifications for new Gage2000H machine.
3.) *With MIPs, the ability to measure a single feature with multiple tips.  (ie a cylinder)
4.) *New result screen:  Coaxial cone and cylinder intersection.
5.) *Tapered probe qualification routine
6.) *Service Tool:  Reload volcomp from software card.
7.) *Alphanumeric headers.  (User can key in company, name, date, etc.) Instructions
8.) *Perpendicular Angles can now be shown as actual angle instead of just gage.
        Instructions
9.) *True Position Polar Coordinates.  Instructions
10.) *Alphanumeric storage program names.  Instructions
11.) XYZ location of  lines is being displayed as centroid of points instead of axis
      intersection point.
12.) Feature recognition engine has improved cone algorithms
13.) Entering in of Qual Sphere diameter value bug fixed (Introduced in 1.07)
14.) Entering in of Squareness limit in DD:MM:SS  bug fixed. (Introduced in 1.07)
15.) Translating to symmetry line and projection point now working correctly.
16.) Don't add 17.0 mm offset to Z-rail unless using scan/done assembly.
17.) Translation to symmetry planes now working correctly.
18.) Softkeys no longer intermittently missing button presses.  (Introduced in 1.07)
19.) Multitip probe qualification from keyed in offsets is now working for tips 2-9.
20.) European language tolerance headers now spaced correctly for printing.
21.) Forced Cone (<2 deg) is no longer showing intermittent 0.000 for intersection circle.
22.) Measurement mode scroll up and down buttons now repeat.
23.) True Position print format variable now being saved.
24.) In Force menu, dim features which require more than 10 pts unless >10 pts taken.
25.) In probe qualification, dim done button until at least 4 points taken.
26.) Relationship between two tapered probe points no longer showing "No Relationship"
27.) European tolerancing ISO table now works correctly in German.
28.) All screens except true position that have diameters now also have radiuses.
29.) Dial Indicator mode now displays runout.
30.) Ability to force a sphere to a plane for slightly spherical planes.
31.) Ability to store the second intersection point of an intersecting circle and plane.
32.) Single point polar coords now shows hole location if approached perp. to Datum A
33.) In Inches, tolerance's default nominal for angle now being displayed correctly.
34.) Dim "Qualify Additional Tip" after 9 tips measured.
35.) Circle, cylinder relationship 2d distance isnow correct after alignment.
36.) Intermittent "Press Done When Finished" printout bug fixed.
37.) Intermittent loss of user defined nominal bug fixed.
38.) Serial out / Print out from Dial indicator mode.
39.) Line x Line angle is now showing actual angle instead of absolute value of angle.
40.) Serial output of measured features cleaned up
41.) True Position RFS now being correctly represented with an S instead of an R.
42.) Improved sound driver such that no "pops" are heard when saving system options.
43.) Separate "Nearest Nominal" variables -inches and mm.  For demo purposes.

 1.09 (3/11/99)

1.)  *Full support for the Gage2000H and Derby MSM machines. ("Move Machine" &
        "Port Features" functions)
2.)  *Support for Mistral machines which use the new TTL controller.
3.)  *Support for Chinese (simplified) Polish, Dutch, & Danish.
4.)  *Ability to delete ANY block.
5.)  *New option:  Require user to take as many points in playback as was orginally
        learned in inspect.
6.)  *With MIPs, the ability to measure a single feature with multiple tips in playback.
7.)  *COM1 now supports the higher19200, 38400, 57600, & 115,200 Baud Rates.
8.)  *COM Port values now stored in EEProm.
9.)  User can no longer set nearest nom to 0.0.  This solves software fault 4x4 problem.
10.)  Per request, user now allowed to set smaller min cyl depth (Inches only - 0.01 in)
11.)  Ability to key in header partname instead of just "Actual" playback part name.
12.)  Compatability with RefleX Command 1.0 Sales/Service tool
13.)  CPK, CR & CP values for internal statistics package fixed.
13.)  ISO tolerancing for non fractional numbers (i.e 10.000 mm) fixed.  Also, changed
        double "ee" to single "e".
14.)  Fixed (printer) data headers for European languages.

 2.00 (7/6/99) (dash 6 controllers don't work with this version)

1.)  *Leitz protocol support for the DCC Microval PFX
2.)  *Support for the Gage2000 A
3.)  *Other Machine Type screen fully supports TTL and Heidenhain controllers.
4.)  *New "Edit Sequence" menu which includes  "Move Last Measured Feature", "Move
        Any Measured Feature", & "Replace Any Measured Feauture"
5.)  *Startup Percentage Bar
6.)  *Improved main menus.  Datum C now shows up on both of relationship menus.
7.)  *New playback "Jump to Below Alignment" softkey for measuring parts in fixtures.
7.)  Bug in "Delete Any Feature" fixed.
8.)  System Options cleaned so that all options are now centered.
9.)  Loss of datum after translation bug fixed.
10.)  Alignment issues from 1.09 Gage2000 A implementation fixed.
11.)  MIP recall tip in playback screens fixed.
12.)  Playback intermittantly dropping Ref. length, tolerances, and concentricity bug fixed.
13.)  System no longer printing extra screen after "Relationship between Any 2 feats".
14.)  Ability to store point created f/ the relationship between concentric cone & cylinder
15.)  Ability to store point & set datum C for relationship between 2 oblique cylinders
16.)  Status bar now correctly showing current probe being qualified during qualification.
17.)  2.00.01 Faro system now working correctly on newer controllers as well as old.

 2.1 (9/24/99) (dash 6 controllers only work with 2.1.01, not 2.1!!!)

1.)  *Unicode Simplified Chinese, Traditional Chinese, Japanese & Korean finished.
2.)  *DCC approch vectors now normalized.
3.)  *DCC demo mode.  (Auto-Repeat DCC program for display purposes)
4.)  Value toleranced to "Out Of Tol" printing even if in tolerance - bug fixed.
5.)  No longer sending out extra data on serial out of true position screens.
6.)  Service Tool: Autotune save on failure working again (2.0)
7.)  User no longer required to reboot system after machine type change for changes to
      take affect.
8.)  2.1.01  New dash 6 / Heidenhain controller distinction methodology.

 2.2 (11/1/99)

1.)  *First User release of DCC.  (Dcc Qualifications added)
2.)  Nominals & Tolerances of 0.000 now being stored correctly for playback.
3.)  Software fault caused by using the demonstration tablet's zmouse (done) fixed.
      (broken in 2.1)
4.)  Done button in playback when "Require Points" disabled fixed.  (2.1)
5.)  Now able to requalify a TTP w/o going back to home screen.  Points now being
      taken correctly in qualify additional tip screen. (2.1)
6.)  Dial Indicator mode printing correctly in Inches.
7.)  Down arrow softkey now appears correctly after scroll up over datum A, B,C.

 2.3 (1/26/00)

1.)  Removed "Configuration - Single / Multi" from Probe Qualification screen.  Qualify       additional tip always available at probe accept screen.  New option:  TTP Probe       Type - TP-ES, or TP-MIP
2.)  DCC:  User now able to continue after dcc error such as "Unexpected Probe Hit"       during qualifications.
3.)  Optical Probes:  User no longer forced to measure qualification sphere if all
      keyed in values are used.
4.)  DCC:  Now approaching correctly after multiple alignments.
5.)  Lower tol. values now defaulting to negative correctly when upper value is 0.000.
6.)  True position on cylinder now working in playback correctly. (2.0)
7.)  Feature ID now always printing out correctly. (2.1)
8.)  Software fault at 104 blocks fixed.
9.)  Fixed cursor problem after relationships in playback.
10.)  Dial indicator displaying max, min, runout properly again.  Print header correct too.
11.)  Horizontal "Other" machine types now supported.
12.)  Statistics now being enabled automatically when a program with stats is run.
13.)  Rare probe not reponding (after reset of measurement mode) bug fixed (2.2?)

 2.4 (6/8/00)

1.)  *New "Pass Through" mode for connecting with PC-DMIS & Xact Measure
2.)  *Work instructions (to screen, printer, serial) now available.(for feature labeling too)
3.)  *Fixturing for parts on non-DCC machines now supported in playback
4.)  *External Vorne display implemented.
5.)  *Abiltity to construct a plane (from the datum) with a specified offset.
6.)  *New Screen:  Coaxial cone cone intersect.
7.)  * "Format Card" is now password protected to prevent accidental deletion of progs.
8.)  Program size increased to 200 blocks.
9.)  Mistral 7.10.5 machine type added.
10.)  Improved sorting methodology for program listings.
11.)  Support for new printers:  Epson LQ570+(US), Epson VP-600 (Japan)
12.)  Fixed graphics during DD:MM:SS input. (2.1)
13.)  Perp distance between a point & line now being displayed correctly for optical prb.
14.)  True Position result values now showing tolerance band correctly.
15.)  Playback translate to last feature fixed
16.)  CPK, CP & CR moved to summary footer.  Also correct for nom=0.000 (2.1)
17.)  Playback with multiple graphic printouts now updating results correctly.
18.)  System operation above 100 blocks cleaned up (1.09)
19.)  Flatness formplots nolonger upside down for YZ & ZX plane(1.06).
20.)  For newer controllers, show print header correctly: Instead of "2/22/0" do "2/22/00"
21.)  On printouts, squareness screens Ref.Angle correctly being shown as 90 in Inches.
22.)  Users no longer allowed to key-in a negative reference length.
23.)  True position polar no longer intermittantly switching to RFS in playback
24.)  Statistics nolonger auto-enabling when screen maked for just printer, serial, etc.(2.3)
25.)  Improved circle-plane feature recognition distinction above 6 points.
26.)  'Coaxial cone-cylinder' & 'plane circle perp. dist' now working correctly in PB.
27.)  Deletion after Relationship between any 2 feats later sequence num. problem fixed.
28.)  Software fault 0x80b... type 4, subtype 4,after upgrade from 1.08 (others?) fixed.
29.)  Full Tolerancing of negative values bug fixed.
30.)  True position working correctly when subsequent "Polar Coords & Dia" marked.

 2.5 (12/17/01)

1.)  *Saved datums available after reset of measurement mode. (but not after shutdown) 2.)  *Added Hungarian (Magyar) text.
3.)  *Added prompt the user to return to the proper probe selection if an abort, cannot
        solve, or a remeasure occured during a multi tip feature measurement.
4.)  *Added fatal memory allocation error screen (Not enough memory to continue)
        that forces the user to restart the system.
5.)  Faro arm TTP qual fixed (2.1)
6.)  Program headers now sent out serial port when appropriate.
7.)  Construct Offset plane after datum now working correcly.
8.)  Slot calculation length now using the current alignment correctly.
9.)  Main measurement menu - punchthrough buttons functionality retired.
10.)  Height gauge zero buttons now only zeroing appropriate axis instead of all 3.
11.)  Qualify Additional tip was reverting to tip 1. Should be tip just qualified.
12.)  Optical probes now display Z value on points for use with microscopes.
13.)  Tip comp (ID/OD) for cyls measured with a <=5 mm probe now working correctly.
14.)  Delete all saved datums when an axis error occurs. (Homing required)
15.)  Probe speed indicator reset if all points are deleted.
16.)  Remove button push through feature.
17.)  Fixed missing Cylinder diameter in screen #1 of ZX plane.
18.)  Fixed graphics of cylinder screens #2 & #3 to display proper plane. (broken in 2.4)
19.)  Init Com2 (Zmouse port) to 9600 baud, 8 data bits, no parity, 1 stop bit, if a mouse         is not selected.
20.)  Remove DCC option.
21.)  Fixed obscure lock up when first aborting from recall probe then aborting back to
        the scan screen. Lock up would occur during next probe hit.
22.)  Correct circle and cylinder true position print out error.
23.)  Fixed cone/plane intersect as datum C error.
24.)  Dont read scan/done buttons if not selected.
25.)  Improve approach point generation.
26.)  Changed Y axis offset for gage 2000H.
27.)  Fixed Epson LX300+ hang up problem.
28.)  Fixed polar/cartesian true position value change bug while stepping back through
        results screens after exiting playback mode.
29.)  Fixed part display scaling of datum screens (different from measurement screens)
30.)  Fixed editing and storing of an executed program causing relationships between         wrong features.
31.)  Changed company name in disclaimer to read Brown & Sharpe Inc.
32.)  Fixed true position stats not always creating a new record for each new feature         learned.
33.)  Fixed blank screen hang up bug.
34.)  Reverted to old (v1.08) memory management scheme to avoid memory allocation
        errors that caused the infamous blank screen hang up bug.
35.)  Force all PCMCIA cards to 5V Vcc if a PD6722 PCMCIA controller chip is
        detected.

 2.5.1 (4/1/02)

      VERSION 2.5.1 IS THE MINIMUM REQUIRMENT FOR THE
     FOLLOWING REFLEX / DERBY ELECTRONICS BOXES AND
     THEIR SUCCESSORS:
                      P/N 180-223-8
                      P/N 195-223-2
                      P/N 198-4016-3
     Older software versions used in these boxes will cause illegible screens.

1.)  Reinstated Direct Computer Control functions that were removed in 2.5
2.)  Fixed stepping beyond either end of the construct a feature list with the Up or Down
       arrows
3.)  Perpendicularity (Squareness) nominal changed from 90 degrees to 0.000 in/mm.
       Result is tilt distance not an angle
4.)  Changed interrupt (7) handeling scheme to improve COM2 response.
5.)  Fixed statistics serial output of part/program and variable names in DataPage format
6.)  Modified block data memory assignment scheme to allow stepping back through
       blocks when the number of blocks is greater than 200
7.)  Modify block data memory management when deleting or moving blocks within a
       program and when resetting the measurement mode.
8.)  When deleting a block other than the last block or when moving a block within a
       program the affected blocks' relationships are updated to reflect their new neighbors.
9.)  Added time delay between writes to LCD controller to allow the use of the SE1335F
       replacement chip.

 2.6 (6/9/03)

1.)  Edge Probe and Optical Probe selections now display X, Y & Z offsets without the
       Diameter while allowing only Type and Diameter for Faro machines
2.)  Fixed recalling Tip #1 in a multi tip feature during Playback
3.)  A printed list of programs stored on a Data Card had more than one name on a line
       if one of the program names was 8 characters long
4.)  Added True Position to MUX10 Stats serial output
5.)  Modified MUX10 Stats serial output format to include a more discriptive letter
       i.e. X for the X axis '01X' Was always an A '01A'
6.)  Corrected stats serial output value for True Position deviation angle
7.)  Modified Stats serial out put of True Position for Generic and GageTalker formats
8.)  Changed Icon in the load from data card screen from DCC icon to the
       'load from data card' icon
9.)  Modified Line From Axis to allow selection of 2D line in a reference plane so it can
       be used in relationships with other 2D features
10.)  Put correct line/circle intersect values in the correct screen i.e. first intersect
        is where line enters the circle not exits it
11.)  Fixed output of internal stats values for CP, CR and CPK broken in 2.4
12.)  Editing (moving blocks and deleting blocks) was not handling stored features
        and reference features. This caused relationships using stored features
        to fail during replay.
13.)  Reflex would hang while trying to list programs. Couldn't find the next
        empty file holder if the card was full.
14.)  Flatness plots in the machine's XY plane were upside down.
15.)  Added measured form error to the minimum approach distance to insure proper
        probe diameter compensation and that inside / outside determinations are correct.
        This also corrected Flatness plot orientation problems.
16.)  Allowed +/- tolerances for 'Actual angles' while leaving the maximum allowed
        type tolerance for 'Gauge angles'. This is also reflected in the printouts.
17.)  Added more resolution to the keyin entry for resquare bar lengths.
18.)  Modified MUX10 to output the same stat values as Data Page.
19.)  Fixed Work Instruction Keyboard screen Abort key function. Use to abort to
        System Options screen. Now it aborts back to the Work Instruction screen.
20.)  *Added Circular Runout screen. Circle screen number 4. It displays
        the circular runout and the Max and Min Radius lengths.

* Denotes enhancements

Please contact RefleXSupport with questions or comments.